Description
- Brand Name: NoEnName_Null
- Origin: Mainland China
- Output: Analog Sensor
- Model Number: 4O2-GP
Basic performance
Working principle: electrochemical two electrodes
Detection range: 0 to 25% o outline dimension diagram
two
Maximum load: 30% O2
Output signal: 100 ± 20 μ A in air
Zero drift (- 20 to 40 ° C) 0.6%
Response time < 15 s
Repeatability < 2% signal value / month
Linearity
Electrical property
Resolution: 0.01% O2
physical property
Weight: 16 g
Orientation requirements: None
work environment
Intended use: suitable for most life safety detectors
Operating temperature range: – 20 to 50 ° C
Operating pressure range: 90 to 110 kPa
Operating humidity range 15% to 90% RH (non condensing)
Life warranty
Recommended storage environment: in a sealed container (0 to 20 ° C)
Service life: 2 years in air
Storage period: 6 months in original packaging
Warranty period: 12 months after delivery