Gold Content and Purity XRF X-ray Fluorescence Spectrometer
Warm Tips:
The price includes the costs of delivery and insurance, which varies in types. For details, please contact our client service.
The Principle of Energy Dispersive X Fluorescence Spectrometer
The atomic electron layer is excited by high-energy X-rays, causing the transition of electrons outside the nucleus, resulting inthe release of secondary X-ray energy.
Because the energy released by different elements has different characteristics, the detection system measures these energies andquantities to calculate element content.
X-MAY72B X-Ray Fluorescence Energy Dispersive Spectrometer Split Type (Equipped with branded computers)
X-MAY72B X-Ray Fluorescence Energy Dispersive Spectrometer Split Type (Not equipped with a computer, the user prepares)
Specification parameters
Gold Content and Purity XRF X-ray Fluorescence Spectrometer
X-MAY72B X-Ray Fluorescence Energy Dispersive Spectrometer Split Type
|
||
Analysis Range
|
0.01% to 99.99%
|
|
Accuracy
|
0.03% (Content≥96%)
|
|
Sample Form
|
Solid, powder, liquid
|
|
Tube Voltage
|
5KV ~ 50KV
|
|
High Voltage Power Supply
|
0 ~ 50KV
|
|
Camera
|
HD camera
|
|
Filters
|
Selectable customized switches
|
|
Detector Type
|
Si-pin (USA)
|
|
Multichannel Analyzer
|
Digital multichannel
|
|
Sample Chamber Size
|
310 *270 *80 mm
|
|
Test Time
|
10sec ~ 100sec
|
|
Measuring Element
|
From sodium (Na) to uranium (U) between all 72 elements
|
|
Analysis Software
|
Qualitative and quantitative analysis software
|
|
External Dimensions
|
460*390*350 mm
|
|
Package size
|
570*540*500 mm
|
|
Weight
|
38 Kg (Spectrometer)
50 kg (Spectrometer and computer)
|
|
Instruments Environmental Requirements
|
||
Ambient Temperature
|
-5 ° C ~ 38° C (Suggested that in air conditioning room)
|
|
Relative Humidity
|
15% ~ 90%
|
|
Power Requirements
|
AC 220V ± 5V, 110V ± 10V , 50/ 60 Hz
No high power electromagnetic and vibration interference sources nearby
|